Receiving and storing 2–8 °C reagents

Most diagnostic reagents lose performance if they get warm or freeze. These habits protect your results — and your money.

01

Before the delivery arrives

  • Keep space free in a lab refrigerator (2–8 °C) for the incoming reagents.
  • Check that the refrigerator thermometer reads between 2 and 8 °C.
  • Tell us your receiving hours so cold boxes are not left waiting.
02

At delivery

  • Open cold boxes immediately and check the ice packs are still cold.
  • Match items, quantities, batch numbers and expiry dates against the invoice.
  • Report damage, a warm box or short supply on the delivery record or to us within 24 hours.
03

Storing reagents

  • Store 2–8 °C reagents in the refrigerator body, not the door and never in the freezer unless the label says so.
  • Keep opened and unopened packs apart and write the opening date on each pack.
  • Use the earliest expiry first (FEFO).
  • Respect on-board stability: some reagents last only a set number of days once opened.
04

Kits and controls

  • Let reagents reach room temperature only if the insert asks for it.
  • Do not mix lots of reagents, calibrators and controls unless the manufacturer allows it.
  • Record control results daily and act on Westgard rule violations.

How we ship cold items: Shipping & cold chain policy.